4. Fitting and Removing CleanSpace2
To carry out a Flow Rate Test:
1.
Fit the yellow Calibration Flow Test Cap to the left bellow
2.
Leave the filter in place and press the ON button to put the
respirator into Standby mode.
3.
Press and release the Flow Test button.
4.
CleanSpace2 automatically runs the Flow Test.
5.
After 2 seconds CleanSpace2 reports the result using the
LEDs on the keypad, as follows:
3 LEDs:
2 LEDs:
1 LED:
3 LEDs and
red filter LED
FLASH:
6. Remove the yellow Calibration Flow Test Cap
Note: The LEDs during the Flow Test will remain lit for 10 seconds, after which CleanSpace2 and the LEDs will
switch off.
STEP 2: ASSEMBLING THE RESPIRATOR
Check that the mask and the Power Unit are correctly assembled and are in the correct orientation. The nose piece
of the mask and the keypad of the power unit should both be aligned. Familiarise yourself with the keypad and the
three (3) modes of operation (see Section on Operating Functions).
To connect the mask to the Power Unit, locate the black clips on the end of the arms of the mask and the
corresponding black clips on the bellows of the Power Unit. To connect, clip one side of the mask and bellows into
place.
Familiarise yourself with the bellow adjustment buttons and check the adjustment on the bellow is at its loosest and
that both bellows are fully extended. Familiarise yourself with the mask release buttons on either side of the bellows.
Leave the other mask arm and bellow free until you are fitting the Respirator.
Attaching the harness
To connect the harness, locate the two (2) straps on the side and one at the back. Hook the clips on the two side
strap contact points on the Power Unit and mask. On the harness, two (2) clips are attached at the sides of the
mask, and the velcro attachment attaches between the neck pad and the centre inside of the Power Unit.
Filter good and battery good (>180 l/min)
Filter ok and battery good
Filter needs changing soon or battery needs charging
System air flow is too low (<120 l/min) to operate. Do not use
CleanSpace2 until a new filter has been fitted and / or battery charged.
If the System airflow is too low, reset the respirator by pressing the Power
button to reset the LEDs to OFF and then fully charge the battery and / or
replace the filter. Repeat the flow test. If CleanSpace2™ fails the test
again, contact CleanSpace Technology and do not use until it has been
evaluated