Fonction
'S'
SET ?
Sends the configuration of the unit for the current channel.
1 or 2 (measuring unit, Resolution, Direction, Rel/Abs, locked
display and keyboard status).
SYS RST
Reset the unit (customer parameters)
STO 0 or OFF
Unlock the display
STO 1 or ON
Lock the display
STO ?
Print out the display status
'U'
UNI ?
Sends the current unit mode (mm or '')
UNI 0 or OFF
Lock the Unit function
UNI 1 or ON
Unlock the Unit function
'V'
VER ?
Sends the version and date of software (Vx.y dd.mm.yy)
RS 232 messages of transmission errors
Code
Display
ERR0
'fct oFF'
ERR1
'rS Err'
ERR2
'rS codE'
ERR4
'rS FULL'
ERR5
'no Func'
ERR6
'rS orun'
ERR7
'rS Err'
ERR8
'rS Err'
ERRA
'MEMO Err'
ERRB
'MEM1 Err'
P1 ERR
'no Prob1'
P2 ERR
'no Prob2'
P12 ERR 'no Prob'
RS 232 parameters of the unit
Baud rate
4800 (transmission speed)
Parity
even
Data Bits
7
Stop bits
2
Flow Control
None
Type of error
Command not executed, deactivated function
Parity error
Unknown format
Capacity Overflow, more than 200 characters without CR
Command not executed, unauthorized function.
Overrun error
Frame error
Stopped transmission error
Not critical error of Flash memory
Critical error of Flash memory, requires a keyboard
acknowledge.
Probe 1 not connected
Probe 2 not connected
Probes 1 and 2 not connected