First part of the diagram, see previous page
....
0.0000
If a channel is not calibrated, the display show :
CAL 1 Err
CAL 2 Err
CAL
Err
If a probe is not connected, the display show :
no Prob 1
no Prob 2
no Prob
By pressing the Preset key when the channel is in absolute mode (Abs) the display
briefly indicates Abs.
3.2 Calibration for diameter measurement
The unit can be calibrated to measure internal diameters.
2 setting rings with different sizes are required to perform the calibration. Use the
smallest ring as Reference 0 and the biggest ring as Reference 1.
The value to insert corresponds to the difference between both diameters.
Once the calibration done, insert the value of the small diameter in the Preset menu.
3.3 Matching of a probe and a unit (option available only on D50S-
PRO)
To increase the measurement's accuracy, i tit possible to match each channel of the
unit with its corresponding probe. This operation is independant of the calibration . It
defines a linear correction curve of maximum 26 points on the probe's measuring
range and specifies a correction for each point. This is to be repeated for both
channels if required.
Acquisition of the reference 1
The display return to the value of channel 1
An error during calibration displays the message
<rEF1 Err>
or
<rEF2 Err> for channel 2
according to the active current function
according to the active current function
48