Example of correction
1) Probe P25 with D50S-PRO without correction —> max error 1.4 um
2) Same instruments, but with correction of 10 points (every 2.5 mm—> max. error
0.7 um
The entry of the points has to comply with the following conditions:
• The numbering of the points must be continuous and must start from the point
number 0 or 1 (if the point 0 is not introduced, it will automatically be created
with a value at zero).
• After each point, the plunger of the probe must always be moved from the
outside to the inside position.
• The correction value is limited to 1mm or 0.01 inch.
The insertion or the modification of the points is to be
made on the channel active at this time, by the RS232-C
Com (See corresponding remote commands COR, LCOR
and NCOR). Once calibrated, the led COR is highlighted.
By simultaneous measurement with 2 probes (Cha 1-2 ou
Cha 1+2) the LED is activated only if both channels have
a correction.
49
E

Werbung

loading