• The difference of correction between 2 consecutive points is limited to the half
of the distance between these 2 points. Example: for 2 points with 1mm
distance from each other, the difference of correction between these 2 points
cannot be bigger than 0.5mm
• The unit must be switched on the channel 1 or 2 (key Fct).
If one of the above condition is not observed, the correction point will not be saved
and the error message Err5 (function not allowed) will be shown.
How to introduce a point by point correction
1) Prepare a set of gauge blocks or an appropriate calibration instrument.
2) Set the unit into the following modes:
High resolution (key Resol)
Positive measuring direction (key +/-)
Relative measuring mode (key Rel/Abs)
3) Set the probe to a reference value and zero it (key Preset). This value
corresponds to the pre-travel of the probe (0.8mm).
4) Set the probe's plunger on the first gauge block or calibration value.
5) Read the displayed value.
6) Enter the first correction value with the remote command COR 1/ +/-x.yyyyy.
The value to enter corresponds to the value of the gauge block less the
readed value on the display (take the sign + or – into account).
7) Repeat the points 4 to 6 for the other guage blocks or calibration values.
8) After the entering of the last point, activate the correction curve with the
remote command COR ON.
4. Test of memory
+
Run a test of memory.
Display FLASH........
..... PASSEd if successful test.
If display show MEM 0 Err,
The unit is still ready to measure correctly but must quickly be
checked by a technical inspection.
If display show MEM 1 Err,
The unit is not ready any more to measure