Endress+Hauser Deltapilot S FMB70 Austauschanleitung Seite 6

Austausch der sensorbaugruppe
Vorschau ausblenden Andere Handbücher für Deltapilot S FMB70:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

KA659P/00/a2/09.06
Assembly of sensor unit with aluminum housing T14/T15
0
2
1
housing T14/T15
(as example T14)
guide
mounting link
on retaining ring
2
1
0
Endress+Hauser
Ex i-devices:
The repair has to be performed such, that the voltage isolation of the Ex ia circuits against ground is maintained.
If required, a test can be performed with 500 Veff over a time period of 60 s.
Ex d-devices:
Attention has to be paid not to damage the threads.
In other case the part has to be exchanged.
O-rings
2
45 69 x 2 62
.
.
EPDM
new
1
retaining ring
3
set screw
4
5
sensor
Any repair of a certified device has to be documented!
This includes stating the device serial number, date of repair, type of repair and name of the repair technician.
Mounting the sensor on the housing
Insert the retaining ring at the sensor head.
1
The retaining ring limits the alignment of the housing on the sensor.
2
Exchange both O-rings at sensor head.
Place housing on sensor head and screw it on the sensor until the edge of
3
the retaining ring (about 5 turns).
4
Align retaining ring with the set screw.
5
Screw housing together with retaining ring that the mounting link will
be positioned in the guide at the housing.
Screw housing on the sensor unit until stop. After appropriate assembly the
housing can be turned for more than a complete turn (380°) until the
retaining ring stops the rotation.
Mounting the instrument
Assembly of the instrument has to be done in the reverse order
to the disassembly, see page 1 (step 5 ... 1 ).
Re-install Deltapilot S, connect cables at terminal.
Adjust housing to desired position.
Important:
After the alignment of the housing tighten the set screw firmly to achieve an
optimal potential equalization between process vessel and ground of the the
housing. Especially the electromagnetic compatibility of the electronics will
be improved.
Calibration
After the assembly has been completed the instrument is operable again.
After the instrument has been powered up, a reset
(general reset) must be entered in the
in field "
reset
".
If a measuring sensor unit with a different range has been installed, the
instrument must be calibrated to the new measuring range according to
the manual.
If the product structure of the instrument changes after the exchange of a
sensor assembly (i.e. a different measuring range), an additional
modification type plate must be installed at the housing.
totally 9 turns
set screw,
(for mounting loosen at half)
groove in housing
screw together until minimum
to retaining ring . . .
. . . screw the housing together with
the retaining ring further on.
Exchange of sensor unit on Deltapilot S FMB70
Code 7864
operation menu
view of housing from below
set screw
retaining ring is positioned in the
lower groove of the housing by screwing
the housing together with the sensor
É
É
operation
guide
2