Agilent Technologies TwisTorr 84 FS PCB 24 Bedienungshandbuch Seite 193

Vdc controller
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 23
The controller can answers with the following response types:
Tab. 7
Type
Logic
Numeric
Alphanumeric
ACK
NACK
Unknown
Window
Data Type Error
Out of Range
Win Disabled
TwisTorr 84 FS PCB 24 Vdc Controller User Manual / 87-901-039-01 (B)
Response type
Length
Value
1 byte
-
6 bytes
-
10 bytes
-
1 byte
(0x6)
1 byte
(0x15)
1 byte
(0x32)
1 byte
(0x33)
1 byte
(0x34)
1 byte
(0x35)
Technical Information
Data Exchange Protocols
Description
Logic value
Numeric value
Alphanumeric value
The command execution has been
successfully completed
The command execu-tion has been
failed
The specified window in the
command is not a valid window
The data type specified in the
command (Logic, Numeric or
Alphanumeric) is not accorded with
the specified Window
The value expressed during a write
command is out of the range value
of the specified window
The specified window is Read Only
or temporarily disabled (for example
you can't write the Soft Start when
the Pump is running)
17
193/210