4.8.3 Assigning the Switching Output for recording

In menu
the switching output to be recorded is assigned. Only one switching output can be
journalized.
Diagnostic
Diagnostics menu
Range:
to

4.8.4 Delay for recording min/max Temperature

A delay time guarantees recording of reliable values for minimum and maximum levels in case of
fluctuating temperature. A time interval (in seconds) is set, during which the signal must at least remain at
maximum or minimum to be recorded.
Diagnostic
Diagnostics menu
Range:
0...100 seconds
Set Jornal Out
Parameter
Set recorded output
Note:
Approx. every three hours the values of the volatile memory are transferred
to the non-volatile memory.
If an alarm is valid when turning the device on or when setting a new output
for recording, this alarm is not recorded.
Delay MM
Parameter
Delay storing Min/Max
Open the parameter menu using
Adjust the value using
pressing
The device returns to sub-menu.
.
and
(e.g. 5 seconds) and accept the value by