Push the spindle of the measuring unit upward by 1 mm or
more, and retract it.
The reference point lamp of the counter module stops flashing
and lights.
3
Align the measuring unit with the master object to be measured.
14
(E)
4
Push in at least 1 mm.
Once this value is set, the reference point will be set automatically
after the power is next turned on when the spindle of the measuring
unit is pushed upward by 1 mm or more. To set the reference point
again, clear the reference point offset value using the RS-232C
command, and proceed again from step 1.

4-2-4. Reset

This function resets all the measured values to zero regardless of
whether the preset value is set.
n
. The reference point setting is also canceled when the reference
point is being used. Set the reference point again.
. If a reset input is supplied while the reference point is being set,
the reference point setting is canceled as soon as the reset input
is supplied.
Input the master value as the preset value using the RS-232C
command.
n
Transmit the SETUP command, P command, then the CLOSE
command, in this order.
The reference point offset value is stored in the counter module.
MG10-P1 / MG10-P2