Maintenance
6.3.2

Warning reference

Warning
Cause
number
#101
Data-logger in Stack mode is 100 % full and
is not logging any more data.
#102
Data-logger in Stack mode is 90 % full.
#103
Data error in the instrument. Some functions
#106
like data-logger, sensor test, etc. may no
longer be available.
#104
Date or time setting not valid.
#107
Battery for data memory empty.
#110
SW dongle has been removed without being
#111
deactivated.
#112
#113
Alarms inhibited.
#164
Sensor warm-up phase has not ended.
#165
Increased measurement error must be
expected.
#167
Calibration interval expired.
#170
#171
Negative reading. Value below range
minimum.
34
Remedy
Download the data, clear the data-logger.
Download the data as soon as possible and clear the data-
logger.
Reset the instrument to factory default settings (see Section 7
on Page 36).
If this error occurs again:
Have the instrument checked by DrägerService.
Set date and time (see Section 5.8.3 on Page 22).
Have the instrument checked by DrägerService.
Deactivate SW dongle (see Section 5.11 on Page 29).
Enable alarms (see Section 5.8.1 on Page 21).
Wait until the sensor has warmed up. Do not calibrate before
sensor is fully warmed-up.
Recalibrate the instrument (see Section 6.2 on Page 30).
Recalibrate zero point.