6075360_BP1700_KAZ Seite 21 Donnerstag, 19. Juli 2007 8:35 08
Problem
Reason
– – –
Inflation above
320 mmHg
While measuring
The device works with
(during deflation
a special deflation
of the cuff),
procedure (stepwise
you can hear a
deflation).
clicking sound
with every
detected
heartbeat.
Readings from
Blood pressure is a
repeated meas-
fluctuating value. For
urements differ
healthy adults, deviations
considerably.
of 10 to 20 mmHg are
possible.
Cuff is not properly
fitted.
Readings were not
consistently taken at
heart level.
Talking, coughing,
laughing, moving etc.
when measuring will
influence the reading.
Solution
Repeat measurement.
none
none
Ensure that the monitor is fitted as
follows:
– on the palm side of your wrist
– approx. 2 cm away from the
palm line
For each measurement, hold the
monitor at heart level.
While measuring, relax, keep still,
do not move or speak.