Alcohol and Soap

Suspension and Lubricant

6
Prepamatic-2
Instruction Manual
Both soap and alcohol are used for cleaning the specimens.
The two containers for these liquids are located in front of the
Recirculation Cooling Unit and accessible through the door in front of
Prepamatic-2.
The container with the blue mark on the lid is to be filled with alcohol
of min. 93% and connected to the blue tube.
The container with the black mark on the lid is to be filled with a
mixture of 100 ml concentrated soap solution, delivered together with
the machine, and 4.9 l of water.
Connect to the black tube.
Connect the correct tube to the proper container.
Seven bottles for suspensions and lubricants are located in the bottle
unit on the left hand side of Prepamatic-2.
Bottles 1 - 6 can be used for both diamond suspension and lubricant,
bottle no. 7 is for oxide polishing suspensions only.
We recommend using bottles 1 - 4 for DP-Suspensions,
e.g. 9µm in bottle 1, 6µm in bottle 2, 3µm in bottle 3 and 1µm in
bottle 4. Bottles 5 and 6 can be used for DP-Lubricant Blue and Red.
Bottle 7 should be filled with either OP-S or OP-U, depending on the
materials to be prepared.
Struers range of DP-Suspensions, lubricants or OP-Suspension are
delivered in bottles that fit conveniently in the bottle unit.
After placing the different consumables in the bottle unit they have to be
entered into the Bottle Configuration in the Configuration Menu.
Please see the section Configuration Menu in the chapter
IMPORTANT
IMPORTANT
Advanced Operations.