2. Attached documents
Manufacturer's declaration - electromagnetic immunity
The CGM MB-1901 is intended for use in the electromagnetic environment specified below.
The customer or the user of CGM MB-1901 should assure that it is used is such an environment.
IEC 60601
Immunity test
test level
Electrostatic
8kV Contact
discharge
15kV Air
(ESD)
IEC 61000-4-2
Electrical fast
2kV for power
Transient/burst
supply lines 1kV
IEC 61000-4-4
for input/output
lines
Surge
1kV differential
IEC 61000-4-5
mode
2kV common
mode
Power frequency
30 A/m
(50 or 60Hz)
Magnetic field
IEC 61000-4-8
Voltage dips, short
0% Uт (100%
Interruptions and
dip in Uт)
Voltage variations
for 0.5 cycle
on power supply
input lines
0% Uт (100% dip
IEC 61000-4-11
in Uт) for 5 cycle
70% Uт (30% dip
in Uт)for 25 cycle
0% Uт (100%
dip in Uт)
for 5 s
Note : Uт is the a.c. mains voltage prior to application of the test level.
48
Electromagnetic
Compliance level
environment - guidance
8kV Contact
Floors should be wood,
15kV Air
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
2kV for power
The main power's quality
supply lines 1kV
should be that of a typical
for input/output
commercial or hospital
lines
environment.
1kV differential
The main power's quality
mode
should be that of a typical
2kV common
commercial or hospital
mode
environment.
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
0% Uт (100%
The main power's quality
dip in Uт)
should be that of a typical
for 0.5 cycle
commercial or hospital
environment. If the user of
0% Uт (100% dip
the CGM MB-1901 requires
in Uт) for 5 cycle
continued operation during
power mains interruptions,
70% Uт (30% dip
it is recommended that the
in Uт)for 25 cycle
CGM MB-1901 be powered
from an uninterruptible
0% Uт (100%
power supply or a battery.
dip in Uт)
for 5 s
2. Attached documents
The CGM MB-1901 is intended for use in the electromagnetic environment specified below.
The customer or the user of CGM MB-1901 should assure that it is used is such an environment.
IEC 60601
Immunity test
Compliance level
test level
Conducted
3 Vrms
3 Vrms
RF
150 kHz to 80
150 kHz to 80
IEC 61000-4-6
MHz
MHz
Radiated
3 V/m
3 V/m
RF
80.0 MHz to 2.7
80.0 MHz to 2.7
IEC 61000-4-3
GHz
GHz
Note 1) Ut is the A.C. mains voltage prior to application of the level.
Note 2) At 80 MHz and 800 MHz, the higher frequency range applies.
Note 3) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should
be considered. If the measured field strength in the location in which the EUT is used exceeds the applicable RF
compliance level above, the EUT should be observed to verifynormal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the EUT.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V 1] V/m.
Electromagnetic
environment - guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the CGM MB-1901, including
cables, than the recommended eparation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
Recommended separation distance distance
Where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters(m).
Field strengths from fixed RF transmitters,
as deter-mined by an electromagnetic site
survey, (a) Should be less than the compliance
level in each frequency range (b).
Interference may occur in the vicinity of
equipment marked with the following symbol:
49