Herunterladen Inhalt Inhalt Diese Seite drucken

Use Of Transmitted Light Illumination - Optika Italy SZR-Serie Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für SZR-Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 74
9.6

Use of transmitted light illumination

1. Set the main switch ① to "I" to turn on the base. (Fig. 12)
2. Turn the light intensity adjustment knob ② to adjust the
brightness to the sample under observation.
3. Move the lever ③ located on the right side of the base. (Fig.
13)
This lever controls a black foil which slides horizontally along
the light exit.
Using the OIC (Oblique Illumination Contrast) system allows
you to modulate the light to obtain three different types of
observation.
4. When the foil is outside the optical path (Fig. 14), brightfield
(BF) observation is obtained.
5. When the foil is completely inside the optical path (Fig. 15),
dark field (DF) observation is obtained.