4
CONNECT THE PURIFIED WATER OUTLET (PORT NO.4)
a. Remove the transit plugs from the water
outlet connection.
N.B This product is not suitable to operate directly with a chorus dispenser.
5
COMMS CONNECTIONS
WARNING!
ONLY USE COMMS CABLE SUPPLIED BY ELGA.
6
CONNECT THE POWER (NO.6 PORT)
3
a. Connect the power cable as shown.
b. Only use the power supply supplied with the Chorus.
The power supply must be earthed.
WARNING!
ONLY USE THE APPLIANCE COUPLER (MAINS LEAD) AND POWER SUPPLY PROVIDED. IF THE EQUIMENT IS USED IN A
MANNER NOT SPECIFIED BY ELGA,THE PROTECTION PROVIDED BY THE EQUIPMENT MAY BE IMPAIRED.
POSITION THE POWER SUPPLY SO THAT IT CANNOT COME INTO CONTACT WITH WATER.
1
2
2
b. Using the tube supplied, firmly push one end
securely into the water connector.
Port 7
or
Port 9
Reservoir J2 or J1
See 'Reservoir
Installation Guide'.
4
PRESS ONCE
c.
to start Commission mode.
Bottom red light will be
flashing until cycle has ended.
Approx 60 minutes