3. MANTENANCE

3.5 Replacing the Composite Vent Filter LC216

Follow the instructions below.

4. SELF HELP GUIDELINES

Constant
Blinking
Constant
Blinking
Press for 10 seconds
Press
(At any time)
Insert a blank USB and
press once. To stop
recording data, press once
(OFF) before removing the
USB.
Set Temperature Alarm
Set Water Purity Alarm
Chorus 2
Over fi ll Protection Alarm
Feature activates drain valve if water level is greater than maximum and stops water entering the
Display Lights
White
White
Red
Red / White
Process Button Operation
Select Languages
Start / Stop Modules
process
Load fi rmware from
Automatically loads system fi rmware from USB fl ash device.
USB Flash devices
Default Alarm Settings
Set default values from fi rmware fi le. Adjustable set point using application
service module. Service module is an engineers tool, only an engineer can adjust the settings.
Select to either 30°C, 35°C, 40°C. Default 35°C. https://elgalab.discovertheweb.co.uk
Set default values from fi rmware fi le. Adjustable set point using application service
module. Resistivity range select from 1.0MΩ.cm,10.0MΩ.cm or 15.0MΩ.cm.
system to prevent a fl ood. When alarm is active:V2 energised. If alarm for 5 mins, stop module
process. Enter refi ll cycle on overfi ll alarm deactivation. Alarm status removed (Reservoir 95%)
Chorus 2 Operator Manual MANU40003 Version 03 3/19 Page 7
Signifying water purity, unit is operating correctly
Consumables are nearing the end of useful life
System needs immediate attention
USB Software update
Language Menu
Start and stops modules process
Re-start on completion.
Default Alarm point 15.0MΩ.cm.