SET - UP
4
CONNECT THE PURIFIED WATER OUTLET (PORT NO.4)
a. Remove the transit plugs from the water
outlet connection.
c. Turn on the water supply.
Minimum Pressure 0.5bar (7.5psi) with boost pump / 4.0bar (58psi) without boost pump.
Maximum Pressure 2.0bar (30psi) with boost pump / 6.0bar (90psi) without boost pump.
N.B This product is not suitable to operate directly with a chorus dispenser.
5
COMMS CONNECTIONS
WARNING!
ONLY USE COMMS CABLE SUPPLIED BY ELGA
6
RESERVOIR NOTE
Water needs to be re-directed from
Port 4a: product outlet to drain during
commissioning mode.
or
b. Using the tube supplied, fi rmly push one end
securely into the water connector.
Port 7
or
Port 9
®
.
O2
or
Connect to:
Reservoir J2 or J1
See 'Reservoir
Installation Guide'.