4.3. Check of flow
•
Switch the Kite Pro on by pressing the ON/OFF button.
After +/- 15 seconds a signal sounds to confirm the
correct function of the electronic control unit.
•
Wait for another 1 minute to be sure that the motor
speed is constant.
•
Place the flow meter on the air outlet in such way that
the flow meter is in vertical position.
•
The indicator ball should be above the minimum level.
If not, the air flow is not sufficient.
•
Warning: placing the flowmeter on the outlet will increase the flow
resistance. The electronic flow control will start compensating by
increasing the flow. Therefore, checking of the flow should not take
more time than 3 seconds.
•
After checking the flow, switch off the Kite Pro.
•
See section 7 for troubleshooting.
•
Warning: Do not close the slits of the flow meter by fingers as
this will influence the correct function.
6