5.5 Cleaning and disinfecting
•
Clean the Kite Pro using a cloth with a mixture of water and a mild detergent. Do not
use any solvents. During cleaning ensure that no water can enter the blower unit. The
blower unit must not be immersed in fluid.
•
Use an alcohol-based disinfecting spray on the inside of the visor and face seal to
remove bacteria and germs. Let it be on the surfaces for 1 minute.
•
Dry the Kite Pro with a cloth and/or leave it to dry in the open air.
•
Warning: do not open the motor unit as all guarantees will be invalidated.
•
Warning: never modify or alter this product as all guarantees will be invalidated.
5.6 Maintenance Kite Pro Vision
5.6.1 Comfort pads
•
Replace all 3 comfort pads when dirty or lost.
5.6.2 Face seal
•
The face seal must be in full contact with the skin
around the face to give maximum protection. Over
time the face seal will lose its elasticity and
therefore it will not function adequately. In doubt
replace the face seal using the following
procedure:
•
Remove the two big black screws on both sides
and remove the visor.
•
Remove one small screw on both sides, see
picture.
•
Remove one small screw on the visor frame
bottom, see picture.
•
Remove the face seal
11