Herunterladen Inhalt Inhalt Diese Seite drucken

KLS Martin marVac smart Vac Gebrauchsanweisung Seite 44

Entrauchungssystem
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 5
iDetect configuration menu
Submenu item
Display
Activation/deact
ivation of the
line current
detector,
section 5.2,
p. 46
Calibration,
section 5.3,
p. 47
(any numerical
value)
Switch-on delay,
section 5.4,
p. 48
(example)
Threshold level
adjustment,
section 5.5,
(example)
p. 51
Flashing
Live display of
the external
power
(any numerical
consumption
value)
44
marVac® / maXium® smart Vac iDetect external line current detector
Description
Line current detection function is
deactivated
Line current detection function is
activated
Line current detector (iDetect) is
not connected
Line current detector is
connected, but a calibration must
follow (teach)
Calibration or measurement
errors during calibration
Power consumption is too low
during calibration
The last measured quiescent
current consumption in mA (or in
A if a decimal point is displayed)
Switch-on delay of the current
detector in seconds
(here: 10 s) at PowerOn
Lower limit of the activation
threshold as a relative reference
to
the measured quiescent current
in percent; if this limit is
undershot, the fan is switched off
(here: 20%)
Active power consumption of the
device connected to the line
current detector in mA (or in % if
a decimal point is displayed)
Instructions for Use
Buttons for changing
parameters
Up button, to activate the line
current detection function and
clear the following submenus
Down button to deactivate the
line current detection function
-
Press and hold the up and
down keys simultaneously for
1 second to start the
calibration process
Up and down button to set the
value within the permissible
limits
Up and down button to set the
value within the permissible
limits
Revision 04