Programming connector on transducer
The programming connector on the transducer is connected
by the programming cable PRKAB 560 to the RS-232 inter-
face on the PC. The electrical insulation between the two is
provided by the programming cable.

Ambient conditions

Nominal range of use
for temperature:
0...15...30...45 °C
(usage group II)
Operating temperature: – 10 to + 55 °C
Storage temperature:
–40 to + 85 °C
Annual mean
≤ 75%
relative humidity:
Altitude:
2000 m max.
Indoor use statement

5.2 Programming the transducer

The transducers SINEAX M 561 / M 562 have an integrated
RS 232 C interface (SCI).
The existing programmation can be matched conveniently
to a changed situation and stored via the "Configuration
software for M 560" (Order number 146 557).
For this purpose, the RS 232 output of the transducer must
be connected to a PC via the RS 232 C (SCI) programming
cable (Order number 147 779 and 143 587) and the trans-
ducer must be supplied with power supply.
The configuration software has an easy-to-operate, clear
menu structure which allows for the following functions to
be performed:
• Reading and displaying the programmed configuration of
the transducer
• Clear presentation of the input and output parameters
• Transmission of changed programmation data to the
transducer and for archiving of a file
• Protection against unauthorized change of the program-
mation by entry of a password
• Configuration of all the usual methods of connection (types
of power system)
Fig. 5. Presentation of all programmation parameters in the main
menu.
28
• Easy change of input and output parameters
WARNING: Watch for maximum input voltage on transdu-
cers with internal power supply connection from measuring
input:
Power supply
Power supply
connection
24 - 60 V AC
Internal from
measuring
85 - 230 V AC
input
• Selection possible for frequency measurement via voltage
or current
• Possibility to reset the slave pointer of the output quantity
involved
• Parameter setting of outputs A and B (input of measu-
red quantity, upper limits, limitation of upper limits and
response time per output, possible up to max. 30 s)
• Graphics display of the set system behaviour of each
output
Fig. 6. Displaying, recording and evaluation of measurements.
Provision is also made for the following ancillary functions:
• Displaying, recording and evaluation of measurements on
a PC
• The simulation of the outputs for test purposes
• Printing of nameplates
6. Reconfiguring the analog outputs
The alternative configurations for the analog outputs can be
seen from Table 1.
Table 1:
Action
Change the current full-
scale value from, for example,
20 mA to 10 mA
(a hardware setting always
thas to be made when
changing from a lower to
a higher value)
Maximum input
voltage across the
power supply
69 V AC
264 V AC
Procedure
Reconfigure the
software,
but do not change the
hardware setting.
Accuracy is reduced.