CL Series Scales
3. OPERATION
3
.1 wo Button Keypad
T
ON/ZERO OFF: Pressing this button turns on the scale. This same button
•
operates the zero feature when the s
three seconds to turn the scale off.
UNIT CAL: Press this button briefly to change the
•
hold this butt
3.2 Calibration
For best results calibrate the scale at regula
th
e scale is in use for prolonged periods.
Cali
bra n weights are not provided with the scale
tio
•
Press and hold the UNIT CAL button to start the
calibration process. The display will show CAL.
•
The calibra
scale off.
•
Press the ON/ZERO OFF button to capture 0. The
display show
load signal.
•
The display will show C xxx w
calibration weight in grams.
•
Place the a
platform.
•
Press the ON/ZERO OFF button.
•
The display shows –C– w
calibration point signal.
•
After calibration, th
weighing mode.
The message CAL E will appear if the cali
followed or the wrong weight was used.
on to begin the calibration process.
tion process can be aborted by turning the
s –C– while the scale stores the zero
ppropriate calibration weight on the
hile the scale stores the
e display returns to the normal
cale is on. Press and hold this button for
r intervals. This is especially important if
here xxx is the
bration steps are not
weighing unit. Press and
EN-3