Herunterladen Diese Seite drucken

Installing And Connecting The Refrigerator - Dometic SILENCIO HC302D Montage- Und Bedienungsanleitung

Medikamentenkühlschrank
Vorschau ausblenden Andere Handbücher für SILENCIO HC302D:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Installing and connecting the refrigerator

If the capacity of the battery is no longer sufficient to take over the monitoring
functions the error message "E3" appears on the display.
A
NOTICE!
The battery should be precautionary replaced every 2 years. This
replacement must be carried out by qualified personnel.
5.5
Internal ventilation
The refrigerator is fitted with an internal ventilation.
The fan is operating when the door is closed.
A
NOTICE!
The airflow to the fan must never be impeded by blocking or covering
the ventilation slots.
To prevent warm outside air from being unnecessarily sucked in, the fan
automatically switches off when the door is opened.
5.6
External alarm function
There are two terminals with two floating contacts on the back of the refrigerator that
can be used to trigger an additional external alarm (acoustical or visual).
6
Installing and connecting the
refrigerator
6.1
Unpacking the refrigerator
1. Check that the packaging is not damaged.
2. Remove the cardboard packaging.
3. Check that the refrigerator is not damaged.
If you find damage on the packaging or at the refrigerator resulting from
transport, report it immediately to the transportation firm.
16
HC302D, HC502D
EN

Quicklinks ausblenden:

Werbung

loading

Diese Anleitung auch für:

Silencio hc502dSilencio hc serie