Changingthesetup
For setup mode, connect the security link and press the
keys simultaneously. The software version will be displayed (flashing). If
you wish to continue, press within 10 s.
Note: If you remove the security link at any time during the setup
sequence, any changes you have made will be discarded and the
instrument will reset.
•Calibration
General
The UCV/AR is factory calibrated. Once calibrated it will operate with
any input or output format without recalibration.
Allow the instrument 1 minutes of powered operation (to reach a
stable temperature) before calibration.
During calibration allow a few seconds for signals to stabilise before
accepting the value.

Equipment requirements

• A suitable current/voltage source (must be able to generate , 10,
20, 0 mA and , 10, 20, 0 V to ±0.0 % or better accuracy)
• An accurate digital multimeter (accurate to 0.0 mV and ±0.1 μA)
TerminalConnectionsforinputcalibration
CalibrationStage
Current (mA) inputs
Voltage (V) inputs
20
Signaltype
Milliamp source +
Milliamp source −
Voltage source +
Voltage source −
Å
and
Terminal
2
3
4
3
ì