Herunterladen Inhalt Inhalt Diese Seite drucken

Gima OXY-6 Betriebs- Und Wartungsanweisungen Seite 136

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 52
All manuals and user guides at all-guides.com
113
Guidance and manufacturer's declaration – electromagnetic emission
The Fingertip Oximeter is intended for use in the electromagnetic environments
specified below, the customer or the user of the Fingertip Oximeter should
assure that it is used in such an environment.
IMMUNITY TEST
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601
TEST LEVEL
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 1 kV line(s)
to line(s)
± 2 kV line(s)
to earth
<5% U
T
(>95% dip in U
)
T
for 0,5 cycle
<40% U
T
(60% dip in U
)
T
for 5 cycles
<70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
3 A/m
COMPLIANCE
ELECTROMAGNETIC
LEVEL
ENVIRONMENT
GUIDANCE
± 6 kV contact
Floors should be wood, con-
± 8 kV air
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
N/A
N/A
N/A
N/A
N/A
N/A
3 A/m
Power
fields should be at levels cha-
racteristic of a typical location
in a typical commercial or ho-
spital environment.
frequency
magnetic

Werbung