Using the M1364A ECG Patient Module to Monitor DECG
Applies to Fetal Scalp Electrode 15133D (not for USA) and 15133E
• NEVER immerse a patient module in water when it is connected to the monitor.
WARNING
• Do not use the M1362A DECG Adapter Cable (or any other adapter cable that connects to fetal
scalp electrodes with exposed wires).
1
Attach a pre-gelled electrode (M1349A) to the DECG Adapter Cable (M1362B).
2
Connect the Fetal Scalp Electrode (15133E/15133D) to the DECG Adapter Cable (M1362B). Position
the Adapter Cable to minimize contact with body fliuds.
3
Peel the backing from the electrode (M1349A) and affix it to the mother's thigh. A good contact
between the electrode and the mother's skin will improve the FHR signal. To get the best signal,
ensure that the skin is clean and dry before attaching the electrode. Ensure that the electrode is correctly
positioned to prevent tension in the cable from pulling on the Fetal Scalp Electrode and injuring
the fetus.