Herunterladen Inhalt Inhalt Diese Seite drucken

Environment Conditions - KaWe EUROLIGHT D30 2.5V Gebrauchsanweisung

Dermatoskope
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 4
Warranty: The warranty for the device totals two
years from the date of purchase with proper handling
and following the operation manual (the warranty
does not cover lamps / rechargeable batteries). LED
lamps are subject to five years of guarantee from the
date of purchase. If any issues arise or repair is necessa-
ry, please contact your dealer.
Warnings: An error may cause extensive heat
posing a threat of burn injury. If the devices heats ex-
cessively, immediately turn the handle off.
Warnings: Do not put the dermatoscope in
fluid and keep fluid off the handle! The dermatoscope
is not designed for surgery, so it is sufficient to clean or
disinfect it with surface alcohol-containing disinfec-
tants. Do not use abrasive cleaning agents! The derma-
toscope shall not be sterilized.
Secondary treatment limitations: Secondary
treatment has a minor effect on these devices. The
service life of the product is usually defined based on
the degree of wear and use-related damage.
Directives
Working area: Clean the surface with a disposable/
paper cloth.

Environment conditions:

t
emperature
o
+10°C to +35°C
peration
s
-25°C to +70°C
torage
t
-25°C to +70°C
ransport
Transportation and storage: Store and transport
dermatoscopes protecting them from dust, moisture
and dirt.
Preparation for cleaning: No special requirements.
Manual cleaning: Clean the dermatoscope external
surface with a damp, soft and lent-free cloth. Do not
put the dermatoscope in fluid and keep body off the
handle! The dermatoscope is not designed for surgery,
so it is sufficient to clean it with surface alcohol-con-
taining detergent. Do not use abrasive cleaning agents!
The device shall not be sterilized. Clean the magnifying
glass with a glass detergent only.
Automatic cleaning: Automatic cleaning is impos-
sible/unnecessary.
Disinfection: Clean the dermatoscope with surface
alcohol-containing disinfectants. Use a slightly damp
cloth.
Sterilization: Scaled lens shall be sterilized in auto-
clave at 134 °C for at least 3 minutes.
Additional information: When sterilizing several
devices in autoclave, do not exceed the maximum load
indicated by the sterilizer manufacturer.
11
r
eLative air humidity
max. 80 %
max. 95 %
max. 95 %

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Eurolight d30 3.5vPiccolight d 2,5v

Inhaltsverzeichnis