Instruction manual Primus 25 advanced®

6.2 Sample program

Select MODE, then PROGRAM and create a new program with NEW. Enter a program name and con-
firm with Enter to access the programming functions.
01
LIDHT
ON
Press Enter to insert a new program step. Move from each parameter to the next with the cursor keys
() or with Enter.
02
TEMP
94.0 °C
0:02:00 +0:00
Enter another program step with Enter. Choose one of the steps listed above by using the () cursor
keys, e.g., LOOP[.
03
LOOP[
30x
Continue by entering a temperature profile.
04
TEMP
94.0 °C
0:00:30 +0:00
05
TEMP
50.0 °C
0:00:30 +0:00
06
TEMP
72.0 °C
0:00:30 +0:05
End the cycling step with LOOP] and with the next step switch the heated lid off.
07
LOOP]
08
LIDHT
OFF
Ramps may be inserted into any program. The ramp starts from the last programmed temperature (here
72 °C) and decreases temperature to the set temperature (here 8 °C) at a constant rate.
09
RAMP
4.0 °C
1.0 °C/s
10
STORE
8.0°C
Forever
11
End
PEQLAB_v0314_E
All manuals and user guides at all-guides.com
120 °C
Heated lid on, temperature 120 °C
+0.0 °C
Denaturation at 94 °C
For 2 minutes
Open a program loop - 30 cycles
+0.0 °C
Denaturation at 94°C
For 30 seconds
(Press 0 to enter the time in seconds)
-0.5 °C
Annealing at 50°C with stepwise
temperature decrement by 0.5 °C (Touch-
down)
+0.0 °C
Elongation with stepwise time increment
by 5 seconds
End the program loop
Lid heat off
Target temperature 8 °C with
a speed of 1 °C per second
+0.0°C
Final temperature 8 °C
+0.00
(Minus key!)
End of program
7