Technical information / Functioning / Parameters
Scaling the measuring
range
(analog output)
Voltage output
0 ... 10 V
Current output
4 ... 20 mA
With the parameter "Lower end of analog output" (ASP) you can
select the measuring value for which the output signal is 4 mA or
0 V. With the parameter "Upper end of analog output" (AEP) you
can select the measuring value for which the output signal is 20 mA or
10 V.
Minimum distance between ASP and AEP = 40% of the final value of
the measuring range.
factory setting
U [V]
10
0
MEW
MEW = value of measuring range
The output signal is between 0 and 10 V in the set measuring range.
It is also indicated:
Flow above the measuring range: output signal > 10 V.
factory setting
I [mA]
20
4
0
MEW
MEW = value of measuring range
The output signal is between 4 and 20 mA in the set measuring range.
It is also indicated:
Flow above the measuring range: output signal > 20 mA.
measuring range scaled
U [V]
10
Q
0
ASP
measuring range scaled
I [mA]
20
4
Q
0
ASP
Q
AEP
MEW
Q
AEP
MEW