8.3
Storage of Measuring Results
8.3.1 Manual Storage
•
A long push of the "store" key immediately stores the current measuring result in
the memory together with a sample number.
•
"HOLD" is displayed.
•
The sample number starts at 1 and is automatically increased by one with each
new stored measuring value.
•
1024 results can be stored.
8.3.2 Automatic Storage
•
A short push of the "store" key activates the automatic storage procedure.
•
"AUTO" is displayed.
•
The instrument automatically checks the temperature stability of the sample. If
temperature is stable within 0.2 degrees for 10 subsequent measurements, then
the value is automatically stored together with a sample number.
•
A minimum of 15 seconds is required per automatic storage procedure.
8.4
Recalling Stored Results
•
A short push of the "Recall/Delete" key shows the latest stored value on the
display for 10 seconds.
•
If the "Recall/Delete" key is pushed again shortly within this time, subsequently
all stored results can be displayed from the latest to the oldest value.
•
If the key is not pushed within 10 seconds the display is switched back to the
current measuring value.
B46IB01
Information:
•
No storage is possible if the "low battery" sign appears on
the display.
•
Activating the "SET" mode deletes all stored results.