Storage of Measuring Results

8.3.1 Manual Storage

A long push of the "store" key immediately stores the current measuring result in
the memory together with a sample number.
"HOLD" is displayed.
The sample number starts at 1 and is automatically increased by one with each
new stored measuring value.
1024 results can be stored.

8.3.2 Automatic Storage

A short push of the "store" key activates the automatic storage procedure.
"AUTO" is displayed.
The instrument automatically checks the temperature stability of the sample. If
temperature is stable within 0.2 degrees for 10 subsequent measurements, then
the value is automatically stored together with a sample number.
A minimum of 15 seconds is required per automatic storage procedure.
8.4

Recalling Stored Results

A short push of the "Recall/Delete" key shows the latest stored value on the
display for 10 seconds.
If the "Recall/Delete" key is pushed again shortly within this time, subsequently
all stored results can be displayed from the latest to the oldest value.
If the key is not pushed within 10 seconds the display is switched back to the
current measuring value.
B46IB01
Hints:
No storage is possible if the "low battery" sign appears on
the display.
Activating the "SET" mode deletes all stored results.
25