10 Readjustment

Proper cleaning of the measuring cell is required before and after each series of
measurements.
Poor cleaning can result in the formation of deposits in the measuring cell. This
can cause offsets of the density readings, and can be recognized by measuring
distilled water.
If the density value deviates more than ±0.001 g/cm
(see Appendix B), then a readjustment may be necessary.
A readjustment is only possible if the density difference between measured and
theoretical value is below 0.100 g/cm
and 25 °C.
How to readjust:
1.
Clean the measuring cell with a suitable cleaning agent until no more residues or
deposits are visible.
2.
Remove the cleaning agent completely from the measuring cell.
3.
Fill distilled water of approx. 20 °C into the measuring cell, carefully avoiding the
introduction of bubbles. The ambient temperature should be between 15 °C and
25 °C.
4.
Turn the instrument off. Turn the instrument on again, pushing both the "ON/
OFF" key and the "Recall/Delete" key at the same time.
5.
The measured density and "ADJ" are displayed.
The water temperature has to be between 15 and 25 °C, the density must be
between 0.900 and 1.100 g/cm
B46IB01
, and the water temperature is between 15
3
.
3
from the theoretical value
3
EN
33