» Now take the arm module off your skin and shake it quickly, with short, sharp movements for at
least 10-20 seconds (at least 6 movements per second). You will hear the seizure alarm go off
» Check whether this is correctly transmitted to and processed by the call system.
Always ensure this product is used solely by people who can diagnose epileptic seizures and who
know how to act in the event of an epileptic seizure.
PLEASE NOTE: This product helps draw attention to clinically urgent seizures. It does not guarantee that
all dangerous seizures will be detected. Users must be aware of this at all times.
PLEASE NOTE: This product cannot be used to establish whether the wearer is having an epileptic
seizure and/or whether a dangerous situation has occurred as a result of an epileptic seizure. In the
event of a NightWatch alert, the caregiver/supervisor must always assess the situation for themselves
and to act accordingly, as instructed by a doctor for example.
6.6

Reading out the NightWatch data

The NightWatch allows you to export the data so that you can review specifi c values measured
by the NightWatch at a later time. This requires the base station to be connected to the internet,
via an Ethernet connection, during monitoring. The NightWatch then automatically searches for
a connection to a designated LivAssured server. If this is successful, the orange LED 2 on the base
station will light up, indicating that a connection has been made. Measurement data will then
be sent (anonymously) from the NightWatch to the server every 5 minutes for as long as the
connection remains active. Please contact LivAssured or visit the website for more information
about this option and how to review the stored data.
PLEASE NOTE: The measurement data read out are indicative only and not intended for medical
diagnostic purposes.
2. This test requires a relatively high amount of movement to set the alarm off because the NightWatch is not yet at its most
sensitive setting (LED 5 is blinking green). If the wearer is lying at rest in bed, its most sensitive setting will be activated (LED 5
continuously green). In this mode, much less movement will be needed to trigger an alarm.
NightWatch
NightWatch user manual v4.0 v3.indd 81
NightWatch user manual v4.0 v3.indd 81
v4.0
2
.
EN
81
15-07-20 16:12
15-07-20 16:12