6

Signals from the arm module and the base station

Both the arm module and the base station display signals about the status of the system. The arm
module does this using a LED on the (white) top. The base station displays signals using the LEDs
on the top and also emits sound signals. The signals used are shown below.
LED 1: Power supply indicator
LED 2: Connected to a server (optional)
EN
LED 3: Not used
LED 4: Arm module connected to the charger
LED 5: All OK, detection active / arm module charge indicator.
LED 6: (Technical) Alarm
A: Arm module indicator light
B: Loudspeaker
The brightness of the indicator on the arm module (A) is set for use in a darkened room and may
consequently be more diffi cult to see in daylight / a brightly lit area.
6.1

Arm module and base station display the same signal

LED 5 and A are lit continuously: All OK - wearer is at rest
The NightWatch can detect seizures at its most sensitive setting. The heart rate is measured
reliably, the connection between arm module and base station is in order, the battery is in order,
the wearer is in a horizontal position (is lying in bed), the armband wearer has been at rest for at
least 2 minutes.