Selecting the measurement mode
Fig. 16: warning when ambient tempera-
ture sensor is plugged in
Lo
Fig. 17: Display after alarm
12.2

E-Mode: Set the emissivity

36
IRT (infrared temperature) – Tamb (temperature ambient)
> 6,5 °C:
Red side flashing
The bar shows red color
Hi is displayed beside the THB icon
IRT – Tamb < - 6,5 °C:
Blue side flashing
The bar shows blue color
Hi is displayed beside the THB icon
IRT – Tamb > 5,75 °C or < -5,75 °C:
or
The buzzer and optional vibration functions will
start at the same time.
According to the following value, "CHK" (check) is
displayed beside the THB mode icon. In this case
please continue to observe.
The bar colour changes according to the Temper-
ature:
4,25 °C < IRT – Tamb < 6,5 °C: red
- 3,5 °C ≤ IRT – Tamb < 4,25 °C: yellow
- 6,5 °C < IRT – Tamb < -3,5 °C: blue
Default emissivity: 0,95
The emissivity can be changed from 0.10 (10E) to
1 (100E).
In E-Mode press Mode key.
Then press Up key or Down key to set the
emissivity.
Press Mode key again to confirm.