Medtronic InterStim iCon Bedienungsanleitung Seite 72

Patienten-programmiergerät
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

Note: Press any arrow on the Navigator key
to clear an information screen.
Screen
Poor
communication
Press
Neurostimulator on
key
72
English 3037
Table 4.3
Information screens
2016-06-01
Description and action
The programmer attempted
to communicate with the
neurostimulator, but
communication was
unsuccessful.
Reposition the
programmer over the
neurostimulator with the
screen facing outward and
try communication again.
If using the detachable
antenna, check that the
antenna is connected
properly, reposition the
antenna, and try
communication again.
You tried increasing
amplitude with the
neurostimulator off.
Turn your neurostimulator
on and try increasing
amplitude again.