An error is
generated and
no operation is
possible (the error
indicator is lit).
* Since any symptoms other than those listed above probably indicate product's failure, contact your dealer or our branch
or sales office.
● In case a LED unit fails, other LED units can still properly irradiate light. However, as the total amount of
light is reduced, the curing time should be extended and curing status should be suffi ciently monitored.
The failed LED unit should be promptly repaired.
■ Never disassemble or modify the unit. Otherwise, heat generation, ignition, or rupture could be caused.
■ In case the unit fails or damages, stop using the unit immediately, turn off the main switch, and pull out the
power plug.
■ Do not attempt to repair the failed unit by yourself but request an expert to do it after providing an
appropriate indication thereon.
EN 26
Downloaded from
www.Manualslib.com
Symptoms
"E01" is displayed
Defective temperature
on the display.
sensor
Disconnection of LEDs
"E02" is displayed
in the upper row or
on the display.
suspected defect of
temperature sensor
"E03" is displayed
Step mode switch
on the display.
remaining pressed
Failure or disconnection
"E04" is displayed
of Violet LED in the
on the display.
upper LED unit
Failure or disconnection
"E05" is displayed
of Violet LED in the
on the display.
midlle LED unit
Failure or disconnection
"E06" is displayed
of Violet LED in the
on the display.
lower LED unit
manuals search engine
Causes
The unit may have detected excessively low or high
temperature, or the temperature sensor may have
failed. Immediately stop using the unit and check it
for any defect.
Use the unit in a temperature range of 10 to 40°C.
In case the trouble is persistent even after turning
on the main switch under the normal temperature
environment, contact your dealer or our branch or
sales office.
An error is being notified after irradiation is
completed, as a temperature fault of LEDs in the
upper row was detected during irradiation in the full
mode.
Check the upper row LEDs upon the occurrence of
this error (see Chapter 6).
The error is reset by turning on the main switch
again.
Foreign material such as dirt may be caught to
cause the switch remaining pressed. Clean the
switch (see Chapter 6).
Turning off the switch releases the error.
The error is reset by turning on the main switch
again.
If the same error occurs every time a Full mode /
Step mode use, the LED unit must be replaced.
Contact your dealer or our branch or sales office.
The error is reset by turning on the main switch
again.
If the same error occurs every time a Full mode use,
the LED unit must be replaced.
Contact your dealer or our branch or sales office.
Note
WARNING
CAUTION
Measures