7

TECHNICAL DATA

Weight
Axiolab 5 microscope stand (depending on version and accessories) ............................. approx. 8 to 20 kg
Ambient conditions
Shipping (in packaging):
Permissible ambient temperature ................................................................................... -40 to +70 °C
Permissible humidity (without condensation) .......................................................... max. 75% at 35 °C
Storage:
Permissible ambient temperature .................................................................................. +10 to +40 °C
Permissible humidity (without condensation) .......................................................... max. 75% at 35 °C

Operation:

Permissible ambient temperature .................................................................................. +10 to +40 °C
Permissible relative humidity (without condensation)............................................... max. 75% at 35 °C
Highest permitted altitude of use .................................................................................... max. 2000 m
Air pressure ........................................................................................................ 800 hPa to 1060 hPa
Degree of pollution .......................................................................................................................... 2
Operational specifications
Operational area ................................................................................................................ Closed rooms
Protective class ....................................................................................................................................... I
Protection type ............................................................................................................................... IP 20
Electrical safety ........................................................... in accordance with DIN EN 61010-1 (IEC 61010-1)
.............................................................................................
Overvoltage category ............................................................................................................................ II
RFI suppression ....................................................................................... complies with EN 55011 Class B
Noise immunity ...................................................................................... complies with DIN EN 61326/A1
Mains voltage for the Axiolab 5 ........................................................................................... 100 to 240 V
Mains frequency ................................................................................................................... 50 to 60 Hz
Power consumption of the Axiolab 5 ............................................................................................100 VA
Fuses in compliance with IEC 127
Axiolab 5 microscope stand .............................................................................. 2 x T 3.15 A/H, 5x20 mm
8
TECHNICAL DATA
430037-7014-111
in compliance with CSA and UL regulations
Axiolab 5