Herunterladen Inhalt Inhalt Diese Seite drucken

Measurement In "Auscultation Mode - Omron HBP-1120 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für HBP-1120:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 90
Caution
• The accuracy of a flashing measurement value that is outside the
measurement range is not guaranteed. Always check the patient's
condition before deciding what steps to take.
NOTE:
• If inflation is insufficient, inflation may restart automatically while
measurement is in progress.
Irregular pulse wave detection function
If the pulse wave interval becomes irregular during measurement, the
irregular pulse wave detection icon will light up.
Body movement detection function
If body movement is detected during measurement, deflation stops for
5 seconds and alternately below the value.
After 5 seconds, measurement resumes, and an attempt is made to
complete measurement in one cycle.
NOTE:
• When the body movement detection function has been activated, the
irregular pulse wave icon appears in the measurement result.
Measurement in "Auscultation Mode"
In "Auscultation Mode", this device does not measure blood
pressure. Measurement should be performed by a health care
professional using a stethoscope.
The healthcare professional uses a stethoscope to determine SYS
and DIA by means of the auscultation method.
1. Make sure the power is on.
"0" is displayed.
2. Press the [Auscultation] button.
The auscultation icon appears and the
device enters "Auscultation Mode".
EN
19

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis