Herunterladen Inhalt Inhalt Diese Seite drucken

smith&nephew RENASYS GO Bedienungsanleitung Seite 39

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 87
Indicator light/
audio:
Status/Alarm
Indicator:
Solid yellow
The light above
the Keypad
Lock button will
illuminate
Solid Cyan
Status/Alarm
Indicator:
Solid Yellow
Status/Alarm
Indicator: Solid
Yellow
The audible alarm
will sound.
The audible
alarm cannot be
paused.
Impact to therapy:
Device is not
delivering therapy.
Device is operating
normally.
Device will stop
delivering therapy.
Device will not
deliver therapy.
39
Steps to remedy:
Either start therapy by
pressing the Play/Pause
Therapy button or power
down the device until therapy
is required.
Device has been locked for
home use.
Contact your healthcare
provider.
Contact your healthcare
provider.