• The motor handpiece can stop automatically on detecting a low battery voltage.
After the motor handpiece is stopped for a short time, it may start again even
though the battery is drained. This is not a fault condition of the battery, it is
a normal characteristic.
• The Lithium-Ion battery is recyclable, but disposal may not be permitted by law
in your location. Return them to your distributor if you are unsure of the proper
disposal method.
• When disposing of the
government for disposal, as it may contain materials which could become
industrial waste.
• When discarding the contra-angle attachment, dispose of it as medical waste.
• Users are responsible for operational control, maintenance and inspection.
• Store the system in a place where the temperature is at -10 to +60 ˚C (14 – 140 ˚F),
humidity at 10 – 85% RH, atmospheric pressure at 500 – 1060 hPa, and the system
is not subject to air with dust, sulfur or salinity.

Adverse Reactions

• There are no known adverse reactions.
A. CLASSIFICATIONS OF EQUIPMENT
• Type of protection against electric shock:
– Class II equipment
• Degree of protection against electric shock:
– Type B applied part
• Method of sterilization or disinfection recommended by the manufacturer:
– See Cleaning, Disinfection, & Sterilization
• Degree of protection against ingress of water as detailed in the current edition of
IEC 529: IPX0
X –SMART
Easy complies to IEC60601-1 safety standard and the requirements
of the CE Marking of Conformity.
X –SMART
Easy, follow the instructions of your local
9