Warning

• Use the specified battery for this product. Never use any other batteries than
those that Dentsply specifies.
• The battery contains an organic solvent that is clear or yellowish in color. Should the
solvent come into contact with your eyes, immediately wash eyes thoroughly with
clean water and contact your doctor. Failure to do so may result in loss of sight.
• Should the solvent come into contact with skin or clothing, immediately wash the
exposed skin thoroughly with clean water and completely wash away the fluid.
Contact your doctor in case of skin irritation.
• The battery contains an organic solvent with vapors that can be vented when the bat
tery is abused, disassembled or overheated. Do not inhale the vapors.
• Do not disassemble or alter the motor handpiece.
• If you should notice a battery fluid leak, deformation of the motor handpiece casing
or partial discoloring, immediately stop use and contact your distributor.
• If the product is not going to be used for a long period of time, disconnect the battery.
• The system may present the possibility of malfunction when used in the presence of
electromagnetic interference (EMI). Do not install the system in the vicinity of any
device that emits EMI.
• Do not use in the presence of flammable products.
• This motor is equipped with an electronic circuit to limit file separation. Nevertheless,
files may still separate due to incorrectly set torque or speed, or the use of fatigued files.
• Dental low voltage electric motors should not be used in the presence of a flammable
anesthetic gas mixed with air, oxygen, or nitrous oxide.
(Note: nitrous oxide by itself is not a flammable anesthetic gas.)
• The contra-angle must be steam sterilized prior to initial use and between patients to
prevent cross contamination. After sterilization, make sure the contra-angle attachment
has reached a temperature below 40 °C (104 °F) before use (see chapter H:
Cleaning,disinfection and sterilization).

Precaution

• This product is intended for use by dental professionals only.
• The appliance must only be used with the manufacturer's original accessories.
• Use the Dentsply charging stand for this product. Never use any other battery
chargers.
• Use only 16:1 ratio contra-angle attachments with this product. Contra-angle
attachments with gear ratios other than 16:1 will cause errors in the display
readings and torque limitation system.
• If the motor stops or runs too fast, discontinue use and call for assistance.
• Do not introduce any oil into the micromotor.
• The system functions normally in the environment where the temperature is at
10 - 40 ˚C (50 - 104 ˚F), humidity at 10 - 85% RH, atmospheric pressure at
500-1060hPa, and with no moisture condensation in the Handpiece or Charging
Stand. Any use outside these limits may cause malfunction.
7