J. TROUBLESHOOTING
When trouble is found, check the following before contacting your distributor. If none
of these are applicable or the trouble is not remedied even after action has been
taken, the product may have failed. Contact your distributor.
X-SMART
EASY Charging Stand
Problem
The power lamp does not turn
on when the charging stand is
plugged into the AC outlet.
Charging lamp light does not
light when handpiece is insert-
ed in charging stand.
X-SMART
EASY Motor Handpiece
Problem
The motor handpiece does not
turn on.
The motor handpiece does not
rotate.
The motor handpiece does not
rotate. ("S.t" is displayed in the
torque box.)
22
Cause
The internal fuse has burnt.
No battery is inserted.
The charging contacts are
dirty.
The battery is defective.
Cause
No battery is inserted.
The battery is depleted.
The battery is defective.
There is an open circuit
inside the motor handpiece.
The contra-angle attachment
is clogged.
There is a short circuit inside
the motor handpiece.
Solution
Contact your distributor.
Insert battery.
Clean the contacts by rubbing
with a cotton swab.
Replace battery.
Solution
Insert battery.
Charge the battery.
Replace the battery.
Contact your distributor.
Clean or replace the contra-
angle attachment.
Contact your distributor.