Sirona Dental Systems GmbH
5
Concluding work
Attach the sheet metal cover and paneling A, referring
to the Service Manual.
Check and adjust the diaphragm 11 as described for di-
aphragm 1 in the Service Manual, "Check and adjust
the X-ray beam for panoramic exposures".
Attach the covers and use the supplied upper cover.
Insert the diaphragm insert in the film holder and swing
diaphragm 1 into place.
Enter the new diaphragm in the unit passport.
Perform test runs with the TSA programs. Make sure
that the procedures run noise-free and collision-free
(actuator sleeve and cover).
Hand the supplied accessories and the operating in-
structions over to the customer.
The unit is now ready.
54 71 607 D3297
D3297.092.02.03.02
5 Concluding work