Programming the diaphragm
4.2
number in service routine 21
The transverse diaphragm must be in the radiation po-
sition, since otherwise no change is possible.
i
NOTE
In the following collimator position 4 is described by
way of example. According to the position of the dia-
phragm to be replaced, other positions can also be re-
ferred to.
Case 1: Previously there was no diaphragm in the
location
Display for diaphragm location 4 (example):
kV 04
mA 00
54 71 607 D3297
D3297.092.02.03.02
1.
1.
2.
3.
4.
X-RAY
S.
2I
04
kV
I I
mA
2.
4
.
Enter the value 11 in the mA display with th
"+" (kV/mA) button.
Press the Memory button.
Press the Return button R.
The patient symbol LED of the corresponding collima-
tor position lights up.
Press the ABV button briefly. The service routine S.21
can be exited.
To exit the service routine switch the unit OFF
and back ON.
4 Select service routine S.21
8
.
S23
3
.
7