Herunterladen Inhalt Inhalt Diese Seite drucken

Caire HELiOS Serie Bedienungsanleitung Seite 10

Vorschau ausblenden Andere Handbücher für HELiOS Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 22
HELiOS 300 & 800
Operating Instructions –
Level Verification
1. Lift the H300/H850 by the contents indicator
strap.
2. Push the bottom backside of the unit forward
until it is straight up and down.
3. Observe the green bar in the contents indicator
window. The window is completely filled with
the green bar when the unit is full of liquid
oxygen.
Portable Full
Checking the H300/H850 contents indicator
10 - ENG
PN 21033264-C4 B | User Manual
Portable Empty
User Troubleshooting
The following information is intended to help you
troubleshoot and solve simple operational problems
that you may experience when using your HELiOS
Portable.
1. The Portable makes a hissing sound.
• Hissing can occur to maintain the correct
operating pressure within the Portable. It is
most likely to hiss after filling or when the
position of the Portable is changed. Hissing
can last for approximately 10 minutes after
filling. Additionally, it can occur when the
flow control valve is at a low setting.
2. The Portable does not pulse after filling.
• If during the filling process, the Portable
is disengaged with the vent valve lever
down, pressure may be reduced within the
Portable causing a delay in the conserving
device function. It may require as much as
60 minutes to restore adequate pressure for
accurate oxygen flow. Additionally, improper
filling or lower than normal operating
pressure in the Reservoir will contribute to
the unit not pulsing.
3. The Portable stops pulsing during use.
• Ensure that both connections of the dual
lumen cannula are firmly attached to the
Oxygen Outlet and Sensor connectors.
• Change the cannula if water droplets are
present from humidified exhaled gas.
Replacing the cannula with a dry one
enhances the ability of the Portable to pulse.
• Ensure that the cannula is not kinked.
• Ensure that the cannula tips remain in your
nostrils and do not slide to one side.
• Ensure that there is oxygen in the Portable.
• With the cannula on, close your mouth and
breathe only through your nose to verify that
the Portable has stopped pulsing.
4. The Portable does not fill.
• Verify that there is oxygen in the Reservoir.
• Ensure that the Portable and Reservoir fill
connectors are fully engaged throughout the
filling process.
5. The Portable vent valve does not close properly
at the end of the filling process.
• If the vent valve fails to close and the hissing
sound and oxygen vapor cloud continue,
carefully remove the Portable by depressing
the release button on the Reservoir. Venting
from the bottom of the Portable will stop in
a few minutes. Allow the unit to warm until
you can close the vent valve. The Portable
may require as much as 60 minutes to restore

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Helios h300Helios h850

Inhaltsverzeichnis