For devices with standby mode, these should be switched on approx. 5 minutes before setting to
avoid incorrect measurements.
If there is an error with the switch-on function, the display "C" will blink and an acoustic signal is
emitted twice. Set the changes again. If necessary, select a higher speed for handpieces.
For detailed information on "Troubleshooting", see the instructions for use which you can down-
load from our website at www.renfert.com/p915.
After-run time of the suction turbine (see Fig. E at the beginning of this
document)
After-run time = the time between switching off a connected consumer and turning off the suction turbine.
There are 6 different times to choose from.
After-run time / sec.
0
3
5
10
20
30
In order to prevent the suction from being switched on / off too often in the case of short interrup-
tions (for example when working with the handpiece), a higher time value can be selected.

Cleaning / Maintenance

The device does not contain any internal parts that require maintenance.
Opening the device other than described in the instructions is not permitted!
Cleaning
To clean the device wipe only with a damp cloth.
Do not use and solvent-based or abrasive cleaning agents.
Empty Dust Drawer
After reaching a set time interval, a request appears to empty the dust drawer:
♦ An acoustic signal sounds 3 x.
♦ Display dust drawer (26) lights up.
♦ The display (24) shows a container filling up.
Before removing the dust drawer, a filter cleaning function should be performed:
⇒ Press Enter key (27) for 2 seconds.
♦ The filter is cleaned (duration approx. 8 sec).
♦ The counter for recording the time interval is reset.
After the filter cleaning has finished:
⇒ Unlock the dust drawer with the clamp fastener (4a).
⇒ Pull the dust drawer (4) forward and empty.
⇒ Replace the dust drawer again until it locks into place.
⇒ Close the clamp fastener (4a).
If the dust drawer has been emptied without prior filter cleaning, the counter must be reset separately to
record the time interval.
⇒ Press the Enter key (27) (counter for recording the time interval is reset).
♦ Acoustic signal acknowledges the information.
♦ Dust drawer display (26) is deleted.
If the dust drawer is not emptied, the dust drawer (26) display will still light up. After switching the
device off and on again, a 3 x acoustic signal is used to indicate the emptying of the dust drawer.
To set the time interval, see Fig. D at the beginning of this document, or refer to the Instructions for Use
on the Internet.
Value in display (24, Fig. 2)
1
2
3
4
5
6
- 5 -
Factory setting