5.
Remove the three allen screws indicated in Figure 12. This will
release the reßector panel. Remove the reßector panel. Be careful
not to scratch or leave Þngerprints on the reßector.
Figure 12: Front screws for securing the reßector panel.
6.
You should now be able to see a long thin light tube, which is
located behind the reßector panel. It is held in place by two sock-
etsÑone at each end of the tube. Grab this tube by each end and
pull it forward out of its sockets.
7.
Mount a new tube into the sockets inside the scanner. Press the
new tube into the sockets by applying pressure at each end of the
tube, never at the middle.
8.
Inspect the reßector for Þngerprints and/or dust. Clean it if neces-
sary.
9.
Replace the reßector panel and its three securing screws.
10. Replace the side panels and their eight securing screws.
11. Close the lower-front panel.
Imacon Flextight Progression UserÕs Guide
Hardware ReferenceÑFlextight Progression
1.
2.
Changing the Light Tubes
3.
35