Herunterladen Inhalt Inhalt Diese Seite drucken

General Description - Velleman DVM854 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 42
Always be careful when working with voltages above 60 VDC or 30 VAC
rms. Keep your fingers behind the probe barriers at all times during
measurement.
Never perform resistance, diode or continuity measurements on live
circuits. Make sure all capacitors in the circuit are depleted.

6. General Description

Refer to the illustration on page 2 of this manual:
1.
Display
3 ½ digits
2.
Function buttons
3.
"V" jack
Insert the red (positive) test lead in this connector to measure voltage
and resistance.
4.
"COM" jack
Insert the black (negative) test lead.
7. Overvoltage/Installation Category
DMMs are categorized depending on the risk and severity of transient
overvoltage that might occur at the point of test. Transients are short-lived
bursts of energy induced in a system, e.g. caused by lightning strike on a
power line.
The existing categories according EN 61010-1 are:
A CAT I-rated meter is suitable for measurements on protected
CAT I
electronic circuits that are not directly connected to mains power, e.g.
electronics circuits, control signals...
A CAT II-rated meter is suitable for measurements in CAT I-
environments and mono-phase appliances that are connected to the
CAT
mains by means of a plug and circuits in a normal domestic
II
environment, provided that the circuit is at least 10 m apart from a
CAT III- or 20 m apart from a CAT IV-environment. E.g. household
appliances, portable tools...
A CAT III-rated meter is suitable for measurements in CAT I- and
CAT II-environments, as well as for measurements on (fixed) mono-
CAT
or poly-phased appliances which are at least 10 m apart from of a
III
CAT IV-environment, and for measurements in or on distribution level
equipment (fuse boxes, lighting circuits, electric ovens).
A CAT IV-rated meter is suitable for measuring in CAT I-, CAT II- and
CAT III-environments as well as on the primary supply level.
CAT
Note that for all measurements on equipment for which the supply
IV
cables run outdoors (either overhead or underground) a CAT IV
meter must be used.
V. 01 – 06/08/2018
DVM854
6
©Velleman nv

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis