ROUTINE MAINTENANCE

Filling the cylinders
The cylinders must be refilled with air in accordance with current regulations.
You may only fill cylinders that:
Comply with the legislation and also have a compliant cylinder valve. The test date and stamp of the approved organisation must
be marked in the cylinder body, and must not have expired.
Do not have any defect liable to give rise to an incident risk (e.g. defective cylinder valve).
Do not show any visible signs of humidity (water drops) at the threaded connection.
The humidity content in breathing air is an important consideration for the proper working of respiratory apparatuses. As a result, make
sure that:
The cylinders contain breathing air complying with standard EN12021, which specifies some data under normal conditions of use,
that is to say atmospheric pressure and ambient temperature.
If they are unintentionally emptied completely, have them completely dried by personnel qualified and authorised by
HONEYWELL.
Cylinder valves are fragile components. Make a visual inspection after each use.
Official inspections by approved bodies:
In accordance with the regulations applicable to pressure devices, compressed-air cylinders for respiratory apparatuses are required to
undergo inspections by an approved body. The body marks the cylinder with the inspection date and its stamp using a durable marking
process.
Cleaning, disinfecting and drying
These operations must only be performed by an agent authorised by HONEYWELL.
After each use:
The breathing apparatus must be cleaned.
The breathing hood must be cleaned and disinfected if necessary.
Cleaning
Firstly, clean the soiled and/or contaminated components with warm water and a sponge, adding a universal cleaning agent (soapy
solution), then wipe with a clean wet sponge.
Check that the cleaning agent chosen does not contain any corrosive components (for example:
organic solvents) likely to attack perishable components and that no liquid or dirt enters cavities
in the apparatus.
Do not introduce a compressed air jet into the cavities containing fragile components such as
diaphragms, springs, valves or seals, so as to not damage them.
Disinfecting
The breathing hood must be disinfected after cleaning and drying, using a disinfectant.
For this purpose we recommend:
Clean and disinfect the breathing hood in an ultrasonic tank:
Immerse the hood into the ultrasonic tank using the product 'EPI U-S' (distributed by HONEYWELL), which is bactericidal,
fungicidal and viricidal, taking care to follow the instructions on the container.
Drying
After cleaning and disinfecting are completed, dry all the parts of the apparatus at a temperature from +5° C to 50° C. Avoid any form of
heat radiation (sunlight, oven or central heating). It is strongly advisable to clean pressure conducting parts (pressure reducer, air
pressure reducing system and coupling sensor) with a low-pressure jet of compressed air in order to eliminate any trace of humidity.
Checks
The working of the apparatus must be inspected every time it is cleaned or if any spare parts are
changed.
The inspection must be carried out an agent authorised by HONEYWELL.
Storage
The clean dry FENZY BIO-S-CAPE can be stored in an appropriate sealed cabinet or cupboard.
Make sure that the apparatus is placed on the support plate and that the strap is not folded.
The FENZY BIO-S-CAPE must be stored in a cool place. The storage area must be dry and free of any gases or fumes. Avoid exposure
to significant direct light or sunlight and keep away from radiant heat sources.
The recommended storage temperature should be between +5° C and +45° C. For special storage conditions ( under outdoor shelters
etc.), please contact our technical department.